SPECIFICATIONS : ZEMIC B3G LOAD CELL
|
Accuracy class |
|
OIML R60 C3 |
Output sensitivity ( = FS ) |
mV/V |
3.0 ± 0.008 |
Maximum capacity ( Emax ) |
t |
0.05/0.1 |
0.25/0.5/1/2.5 |
5/7.5/10 |
Maximum number of load cell intervals ( nLC ) |
|
3000 |
Ratio of minimum LC verification interval Y = Emax / vmin |
|
7000 |
10000 |
12000 |
Combined Error |
%FS |
≤ ± 0.0200 |
Minimum dead load |
kg |
0 |
Safe overload |
of Emax |
150 % |
Ultimate overload |
of Emax |
300 % |
Zero balance |
of FS |
≤ ± 1.0 % |
Excitation, recommended voltage |
V |
5 ~ 12 |
Excitation maximum |
V |
18 |
Terminal resistance, input |
Ω |
430 ± 50 |
Terminal resistance, output |
Ω |
351 ± 2 |
Insulation impedance |
MΩ |
≥5000 ( at 50VDC ) |
Temperature range, compensated |
°C |
-10 ~+40 |
Temperature range, operating |
°C |
-35 ~ +65 |
Element material |
|
Stainless steel |
Ingress Protection (according to EN 60529) |
|
IP67 |
ATEX classification (optional) |
|
II1G Ex ia II1C T4 |
II1D Ex iaD 20 T73°C |
II3G nL IIC T4 |